Nitrogen Washout Procedure Overview
Apr 17, 2025
Nitrogen Washout Procedure
Calibration
Connect Umbilical
Connect to the gas calibration port.
Select "gas calibration."
Wait for the "gas calibration successful" message to appear at the bottom of the screen.
Pre-Test Preparations
Connect Umbilical
Connect to the pre-even flow sensor.
Patient Setup
Instruct the patient to place lips tightly around the mouthpiece.
Apply a nose clip to the patient.
Test Execution
Beginning the Test
Select "start" for the patient to begin breathing normally.
Observe four countdown breaths in the quality control gauge.
Perform four tidal breaths visible on the screen.
Breathing Instructions
Instruct the patient to exhale fully at the bottom of the fifth tidal breath and then inhale maximally.
Patient should return to normal breathing.
Select the space bar at the end of the tidal breath.
Shutter will switch the patient to breathing 100% oxygen.
Monitoring During Test
Quality Control Gauge
Displays respiratory rate, accumulated FRC, percent of nitrogen, and switching error.
Ensure switch-in occurs within ±250 mls.
End of Test
Automatic Stop
The test stops automatically when the final N2 percent is ≤ 1.5%.
Confirmation will be shown on the QC gauge.
Results
Scorecard
Displays which ATS criteria were met.
📄