🔬

Quality Control for Thoracic Gas Volume

Apr 17, 2025

Quality Control Test for Measuring Thoracic Gas Volume

Introduction

  • Quality control is crucial to ensure system functionality.
  • This guide demonstrates performing a quality-control test using an isothermal bottle.

Equipment & Setup

  1. Isothermal Bottle Connection

    • Remove the screw on the bulkhead cover and open it.
    • Disconnect the umbilical connector from the purple Gore fitting (mouth port).
    • Attach the tube from the isothermal bottle to this port.
  2. Test Subject Setup

    • Add a new patient for the test.
    • Ensure BTPS correction is unchecked for syringe mode testing.
    • Enter subject's height and weight; add 12 pounds to account for bottle displacement.
  3. Testing Environment

    • Navigate to the "Plath" testing screen.
    • Place the subject in the box with the isothermal bottle and seal the door.
    • Wait 30 seconds for box calibration.
    • Stabilize the bottle using a book to prevent warming.

Testing Procedure

  1. Tidal Breathing & Shutter Operation

    • Click "TGV Start" and have the subject perform tidal breathing with the flow sensor.
    • No need to attach the flow sensor to the patient's head.
    • After 4 stable breaths, press spacebar to close the shutter.
  2. Syringe Pumping

    • Subject pumps syringe at a rate of once per second.
    • Subject holds breath when shutter is closed.
  3. Repeating the Test

    • Press spacebar again to repeat the procedure.
    • Use wrist action for syringe pumping to avoid artifacts.
    • Complete 4 to 5 effort cycles.

Results Review

  • Examine each effort to verify angles are correct.
  • Adjust line of best fit as necessary.
  • Select several acceptable efforts to average.
  • The average TGV values should be within ±5% of the bottle's labeled volume.

Troubleshooting

  • If TGV values are not within range, contact technical support at 1-800-333-4137.