Lecture on the STERRAD 100NX Sterilization System with All Clear Technology

Jul 11, 2024

Lecture on the STERRAD 100NX Sterilization System with All Clear Technology

Overview

  • STERRAD 100NX System: General-purpose low-temperature sterilizer.
  • Technology: Uses hydrogen peroxide vapor and low-temperature plasma for sterilization.
  • Benefits: Rapid sterilization without toxic residue, suitable for heat and moisture-sensitive instruments.

Pre-cycle Diagnostics and Load Conditioning

  • Pre-cycle Diagnostics: Automatically run before each cycle to detect and correct issues.
  • Load Conditioning: Takes about 5 minutes; attempts to correct load issues before starting the cycle.
  • Notifications: Displays if issues cannot be corrected.

Available Sterilization Cycles

  • Standard Cycle (47 min): For general metal/nonmetal instruments, hinged devices, and single-channel stainless steel lumens.
    • Example instruments: Arthroscopy, laparoscopic instruments, cameras, light cords.
  • Flex Cycle (42 min): For single-channel flexible endoscopes with specific dimensions.
    • Max two flexible endoscopes per load.
  • Duo Cycle (60 min): For medical devices and flexible endoscopes with accessories.
    • Max two flexible endoscopes per load.
  • Express Cycle (24 min): For general medical devices requiring surface sterilization.
    • Example instruments: da Vinci 3d endoscopes.
    • Note: Always check the medical device manufacturer’s instructions.

Cassette Handling and Insertion

  • Cassette Requirement: Display notification for new cassette when needed.
  • Handling Precautions: Use chemical-resistant gloves.
  • Insertion Steps:
    1. Take a new cassette, check expiration and moisture.
    2. Insert correctly following arrow direction.
    3. RFID label controls usage and prevents cassettes from being used post-shelf life.
    4. Disposal: Used cassettes moved to disposal box automatically.

Load Preparation and Reprocessing

  • Device Preparation: Clean, rinse, and dry devices thoroughly.
  • Tray Recommendations: Use STERRAD instrument accessories and Optimax trays.
    • Avoid materials like linen or foam pads.
    • Use polypropylene sterilization wrap and Tyvek pouches.
  • Indicators: Use STERRAD chemical indicators for verification.

Loading the Sterilizer

  • Guidelines:
    • Ensure proper diffusion of H₂O₂.
    • Arrange items to avoid blocking UV lamp.
    • Maintain space between load and electrode.
  • Cycle Initiation: Select cycle and press start; system provides loading instructions.
    • All Clear Technology: Detects and corrects problems before starting sterilization.

Cycle Completion and Cancellation

  • Completion: Indicated by green screen, beep sound, screen status “process complete.”
  • Verification: Confirm chemical indicator color change, cycle printout parameters.
  • Cancellation: Touch cancel button; ensure rewrapping and reprocess.

Biological Indicator (BI) Processing

  • Routine Monitoring: Recommended for every cycle.
  • CycleSure 24 BI:
    1. Ensure room temperature and purple growth medium.
    2. Label and place in Tyvek pouch.
    3. Place in sterilizer and run desired cycle.
    4. Post-cycle, check for color change.
  • Velocity BI System:
    1. Prepare BI as per instructions.
    2. Scan barcode and place BI in sterilizer.
    3. Follow post-cycle procedures as per instructions.

General Maintenance and Utilities

  • Power Switch: Located on the left side; system auto-sleeps when not in use.
  • Printer Paper Replacement:
    • Located on front panel; follow steps to replace.
  • Lens Cleaning:
    • Clean every three months or when debris is present.
    • Use isopropyl alcohol and chemical-resistant gloves.
  • Peroxide Clearance: Required for Duo cycles; follow notification prompts.
  • Planned Maintenance: Based on time/cycle count; contact ASP help desk.
  • Cycle History: Stores last 200 cycles; must download history when memory is full.

Conclusion

  • User Guide: Refer to the user guide for complete equipment information, warnings, and instructions.