Lecture on the STERRAD 100NX Sterilization System with All Clear Technology
Jul 11, 2024
Lecture on the STERRAD 100NX Sterilization System with All Clear Technology
Overview
STERRAD 100NX System
: General-purpose low-temperature sterilizer.
Technology
: Uses hydrogen peroxide vapor and low-temperature plasma for sterilization.
Benefits
: Rapid sterilization without toxic residue, suitable for heat and moisture-sensitive instruments.
Pre-cycle Diagnostics and Load Conditioning
Pre-cycle Diagnostics
: Automatically run before each cycle to detect and correct issues.
Load Conditioning
: Takes about 5 minutes; attempts to correct load issues before starting the cycle.
Notifications
: Displays if issues cannot be corrected.
Available Sterilization Cycles
Standard Cycle (47 min)
: For general metal/nonmetal instruments, hinged devices, and single-channel stainless steel lumens.
Example instruments: Arthroscopy, laparoscopic instruments, cameras, light cords.
Flex Cycle (42 min)
: For single-channel flexible endoscopes with specific dimensions.
Max two flexible endoscopes per load.
Duo Cycle (60 min)
: For medical devices and flexible endoscopes with accessories.
Max two flexible endoscopes per load.
Express Cycle (24 min)
: For general medical devices requiring surface sterilization.
Example instruments: da Vinci 3d endoscopes.
Note
: Always check the medical device manufacturer’s instructions.
Cassette Handling and Insertion
Cassette Requirement
: Display notification for new cassette when needed.
Handling Precautions
: Use chemical-resistant gloves.
Insertion Steps
:
Take a new cassette, check expiration and moisture.
Insert correctly following arrow direction.
RFID label controls usage and prevents cassettes from being used post-shelf life.
Disposal: Used cassettes moved to disposal box automatically.
Load Preparation and Reprocessing
Device Preparation
: Clean, rinse, and dry devices thoroughly.
Tray Recommendations
: Use STERRAD instrument accessories and Optimax trays.
Avoid materials like linen or foam pads.
Use polypropylene sterilization wrap and Tyvek pouches.
Indicators
: Use STERRAD chemical indicators for verification.
Loading the Sterilizer
Guidelines
:
Ensure proper diffusion of H₂O₂.
Arrange items to avoid blocking UV lamp.
Maintain space between load and electrode.
Cycle Initiation
: Select cycle and press start; system provides loading instructions.
All Clear Technology
: Detects and corrects problems before starting sterilization.
Cycle Completion and Cancellation
Completion
: Indicated by green screen, beep sound, screen status “process complete.”
Verification
: Confirm chemical indicator color change, cycle printout parameters.
Cancellation
: Touch cancel button; ensure rewrapping and reprocess.
Biological Indicator (BI) Processing
Routine Monitoring
: Recommended for every cycle.
CycleSure 24 BI
:
Ensure room temperature and purple growth medium.
Label and place in Tyvek pouch.
Place in sterilizer and run desired cycle.
Post-cycle, check for color change.
Velocity BI System
:
Prepare BI as per instructions.
Scan barcode and place BI in sterilizer.
Follow post-cycle procedures as per instructions.
General Maintenance and Utilities
Power Switch
: Located on the left side; system auto-sleeps when not in use.
Printer Paper Replacement
:
Located on front panel; follow steps to replace.
Lens Cleaning
:
Clean every three months or when debris is present.
Use isopropyl alcohol and chemical-resistant gloves.
Peroxide Clearance
: Required for Duo cycles; follow notification prompts.
Planned Maintenance
: Based on time/cycle count; contact ASP help desk.
Cycle History
: Stores last 200 cycles; must download history when memory is full.
Conclusion
User Guide
: Refer to the user guide for complete equipment information, warnings, and instructions.
📄