DLCO Quality Control Steps for Testing
Apr 17, 2025
Notes on DLCO Quality Control for MGC Diagnostics Pulmonary Function Systems
Introduction
Quality control (QC) for DLCO on MGC Diagnostics systems using Asent software.
Recommended weekly per ATS guidelines.
Perform DLCO test with BTPS correction turned off.
DLCO Testing Procedure
Preparation:
Add DLCO test to the same patient visit as syringe spirometry testing.
Proceed to DLCO test screen.
Calibration:
Place the umbilical clip into the calibration port.
Perform a DLCO gas calibration.
Once passed, connect the umbilical clip to the prevent flow sensor.
QC Setup:
Use the white DLCO QC coupler (smaller internal diameter for better gas mixing).
Place coupler between the prevent flow sensor and the syringe.
Connect the syringe and flow sensor to the system head.
Testing Steps:
Pull syringe plunger to about 1 L mark.
Initiate test: Click start or press space bar.
Perform 3-4 tidal breaths, then push plunger to 2 L mark.
Press space bar to close shutter and pull back 2 L of diffusion gas.
System locks for 20 seconds for gas mixing.
Push plunger fully in after 20 seconds to stop effort.
Verification and Results
Check data from efforts:
IVC Range:
Should be between 1.90 to 2.10 L.
DLCO Measurement:
Must be < 0.5 mL/min/mmHg or < 0.166 moles/min/kPa (SI units).
VA Verification:
Use MGC QC table to find target VA and verify it's within specified range.
Recalibration:
If IVC is not within range, recalibrate flow sensor and repeat.
If VA or DLCO are not within range, perform another gas analyzer calibration.
ATS ERS DLCO Guidelines (2017)
Weekly DLCO QC check or when issues are suspected.
Perform in normal patient test mode (BTPS correction ON).
From Ascent home screen, select "Add Patient" and enable BTPS correction.
Enter only ID and name.
Testing:
Use fixed anatomical dead space setting (value printed on syringe).
Perform DLCO gas calibration before each effort.
Use QC coupler with syringe and flow sensor setup.
For tidal breaths, push plunger fully in and perform 3 L diffusion gas pull.
Target VA calculated using syringe volume and correction factor.
VA must be within 300 mL of target.
Measurements and Specifications
Room Conditions Example:
Temperature: 72°F (295 K), Pressure: 740 mmHg.
Example Target VA: 3.26 L.
Specification:
DLCO measurement must meet predefined limits.
Recalibration:
Recalibrate flow sensor and repeat if measurements are off.
Settings Reminder:
Restore DLCO F5 option screen settings if changed.
Additional Quality Control
Perform biological QC regularly.
Use healthy, non-smoking subjects.
Reference ATS pulmonary function lab manual for details.
Further Information
MGC Diagnostics product bulletin part number: 1 14255 D001.
Contact MGC Diagnostics technical support at 800-333-4137 or email
[email protected]
.
📄