PRP Preparation Procedure Summary- 30 ML

Nov 16, 2025

Overview

This transcript outlines the step-by-step procedure to prepare platelet-rich plasma using the Pure 2 platelet concentrating system, including setup, centrifugation, and PRP collection.

Materials and Preparation

  • Pure 2 separator device and counterbalance device
  • Sterile syringes: 30 mL and 12 mL
  • Citrate anticoagulant: 3 mL per 30 mL blood draw
  • Sterile alcohol swabs; clear non-vented caps
  • Centrifuge with rotor and Pure 2 IFU protocol settings

Procedure Summary Table

StepActionDetailsKey Checks
1Prepare separator portRemove Red Cap; swab with sterile alcohol before and after accessMaintain sterility
2Prepare syringeFill 30 mL syringe with 3 mL citrate anticoagulantCorrect anticoagulant volume
3Draw bloodDraw 27 mL patient blood to reach 30 mL totalGentle handling
4Load separatorAttach 30 mL syringe to needleless port; slowly inject anticoagulated bloodSlow injection
5CounterbalanceUse counterbalance device with equal volumeEqual volume on both sides
6First centrifugationPlace devices opposite in rotor; follow Pure 2 IFU protocol; startSecure lid; correct settings
7Aspirate PPSAttach sterile syringe to top port; aspirate platelet plasma suspensionPPS above red blood cells
8Transfer to concentratorHold concentrator upright; transfer PPS through bottom portSteady transfer
9Cap before spinAttach clear non-vented cap to bottom port; do same for counterbalance deviceMandatory step
10Second centrifugationPlace devices opposite; follow IFU settings; startEqual volume; secure caps
11Remove excess plasmaAttach syringe to needleless port; aspirate plasma; leave 3โ€“4 mL or desired amountBuffy coat visible at bottom
12Resuspend plateletsRock and swirl until Buffy coat fully resuspends into plasmaNo visible Buffy coat remains
13Collect PRPAttach 12 mL syringe to needleless port; tilt device; aspirate PRPSmooth aspiration

Detailed Steps

  • Port preparation and access
    • Remove Red Cap from needleless valve port of separator.
    • Swab port with sterile alcohol before and after syringe access.
  • Anticoagulation and blood draw
    • Fill 30 mL sterile syringe with 3 mL citrate anticoagulant.
    • Draw 27 mL blood to reach 30 mL total volume in the syringe.
  • Loading and first centrifugation
    • Attach 30 mL blood syringe to separator needleless port; inject slowly.
    • Counterbalance using equal volume device; place opposite in rotor.
    • Follow Pure 2 IFU protocol settings; close lid; press start.
    • After spin, platelet plasma suspension is above red blood cell layer.
  • Transfer to concentrator and mandatory capping
    • Attach sterile syringe to top port; aspirate platelet plasma suspension.
    • Hold concentrator upright; transfer suspension through bottom port.
    • Attach clear non-vented cap to bottom port before centrifugation.
    • Cap bottom port of counterbalance device as well.
  • Second centrifugation and concentration
    • Counterbalance with equal volume; place opposite in rotor.
    • Follow Pure 2 IFU protocol settings; close lid; start centrifuge.
    • After spin, platelet Buffy coat is visible at device bottom.
  • Plasma removal and resuspension
    • Attach sterile syringe to needleless port; aspirate plasma.
    • Leave 3โ€“4 mL plasma or desired amount in device.
    • Rock and swirl until all platelets are resuspended into plasma.
    • Ensure no Buffy coat remains visible at the bottom.
  • PRP collection
    • Attach 12 mL syringe to needleless port.
    • Tilt device to aspirate platelet-rich plasma smoothly.

Key Terms & Definitions

  • Platelet plasma suspension (PPS): Plasma layer containing platelets above the red blood cell layer after the first spin.
  • Buffy coat: Concentrated layer rich in platelets located at the bottom after the second spin.
  • PRP (Platelet-Rich Plasma): Final platelet-enriched plasma collected after resuspension.
  • IFU: Instructions for Use; protocol settings specific to the Pure 2 system.

Action Items / Next Steps

  • Verify centrifuge protocol settings per Pure 2 IFU before each spin.
  • Always attach clear non-vented cap to bottom port before the second spin.
  • Ensure equal volume counterbalance placement at opposite rotor positions.
  • Confirm complete resuspension of Buffy coat before PRP aspiration.