Guide to Alaris PCA System Setup
Feb 13, 2025
Lecture Notes on Setting Up Alaris PCA System
Introduction to Alaris PCA System
Purpose
: Administrating primary fluids and PCA analgesia
Components
:
Alaris channel for primary fluids
Alaris key for PCA
PCA channel for syringe insertion
Monoject syringe from pharmacy (color-coded blue)
Preparing the PCA Setup
Initial Setup
Primary Fluids Setup
:
Primed with LR (Lactated Ringer's solution)
Ready through primary tubing
Preparing PCA Syringe
Monoject Syringe
:
Connect to the PCA syringe channel
Insert into PCA barrel by twisting
PCA Tubing
Attachment
:
Connect Monoject syringe to long end of PCA tubing
Ensure all PCA tubing clamps are engaged
Priming and Programming PCA
Priming Process
Connect Syringe to Tubing
Turn on the Pump
Open Plastic Case
:
Use key from Pixus, turn counterclockwise
Insert syringe, aligning under the lip
Secure with plastic piece, lower clamps, and secure barrel
Prime Tubing
:
Go to options, select 'prime set with syringe'
Unclamp tubing
Press hold to prime, stop automatically at 2 ml
Adjust if necessary to reach the Y port
Programming Infusion
Select Options
:
Confirm time
Select syringe type (Monoject 35)
Confirm drug (morphine)
Set PCA under specific order (e.g., under 65)
Confirm dosage: 5 mg/ml
Require second RN sign-off
Finalizing Infusion Setup
Priming PCA and Primary Fluids
Additional Priming
:
Attach primary fluids to another port
Release clamps to prime PCA tubing by gravity
Program Infusions
PCA Dosage
:
Dose: 1 mg
Lockout Interval: 10 minutes
Max Limit: 15 doses
Concentration: 5 mg/ml
Primary Fluids
:
Set basic infusion rate: 125 of 950 to be infused
Start Infusion
Finalize Setup
:
Ensure all doors are closed and locked
Press start for PCA and primary fluids infusion
Conclusion
Operational Checks
:
Ensure patient uses green button for PCA delivery
Return key to appropriate location
📄
Full transcript