Burette Calibration Procedure Overview

Sep 5, 2024

Calibration of a Burette

Equipment Needed

  • Balance
  • Funnel
  • Distilled Water
  • Burette Holder
  • Burette
  • Waste Water Beaker
  • Small Glass Beaker (for recordings)
  • Larger Glass Beaker
  • Thermometer

Steps for Calibration

  1. Preparation
    • Fill the larger beaker with distilled water and insert the thermometer.
    • Allow to sit for at least 10 minutes to reach room temperature.
    • Record the temperature from the thermometer.
  2. Burette Preparation
    • Rinse the burette with distilled water and let it dry completely.
    • Check that the stopper works and there are no leaks.
  3. Filling the Burette
    • Use the funnel to fill the burette with distilled water, ensuring no air holes are present.
    • Fill it past the zero mark, then remove the funnel.
    • Slowly release water until the meniscus sits on the zero mark.
    • Wait for 5 minutes to check for air bubbles and leaks.
  4. Mass Recording
    • Place a small glass beaker under the burette and record its mass.
    • After confirming the meniscus is still at zero after 5 minutes, release the water until it reaches the 50 cm³ mark.
  5. Final Measurements
    • Record the mass of the beaker and water.
    • Repeat this process three times to gather data for three trials.

Calculations

  • Mass of Water
    • Subtract the mass of the empty beaker from the mass of the beaker with water for each trial.
    • Calculate the mean mass of water from the three trials:
      • Mean Mass = (Mass1 + Mass2 + Mass3) / 3
      • Result: Mean Mass of Water = 49.867 grams.
  • Volume Calculation
    • Use the equation:
      • Volume = Mass × Z-value
    • Z-value: Conversion factor based on water density at the measured temperature.
    • Since the water temperature was 19°C, Z-value = 1.0027.
    • Volume in Burette = 49.867 grams × 1.0027 = 50.0016 cm³.

Conclusion

  • Calibration of the burette is complete with a calculated volume of 50.0016 cm³.