Burette Calibration Procedure Overview
Sep 5, 2024
Calibration of a Burette
Equipment Needed
Balance
Funnel
Distilled Water
Burette Holder
Burette
Waste Water Beaker
Small Glass Beaker (for recordings)
Larger Glass Beaker
Thermometer
Steps for Calibration
Preparation
Fill the larger beaker with distilled water and insert the thermometer.
Allow to sit for at least 10 minutes to reach room temperature.
Record the temperature from the thermometer.
Burette Preparation
Rinse the burette with distilled water and let it dry completely.
Check that the stopper works and there are no leaks.
Filling the Burette
Use the funnel to fill the burette with distilled water, ensuring no air holes are present.
Fill it past the zero mark, then remove the funnel.
Slowly release water until the meniscus sits on the zero mark.
Wait for 5 minutes to check for air bubbles and leaks.
Mass Recording
Place a small glass beaker under the burette and record its mass.
After confirming the meniscus is still at zero after 5 minutes, release the water until it reaches the 50 cm³ mark.
Final Measurements
Record the mass of the beaker and water.
Repeat this process three times to gather data for three trials.
Calculations
Mass of Water
Subtract the mass of the empty beaker from the mass of the beaker with water for each trial.
Calculate the mean mass of water from the three trials:
Mean Mass = (Mass1 + Mass2 + Mass3) / 3
Result: Mean Mass of Water = 49.867 grams.
Volume Calculation
Use the equation:
Volume = Mass × Z-value
Z-value
: Conversion factor based on water density at the measured temperature.
Since the water temperature was 19°C, Z-value = 1.0027.
Volume in Burette = 49.867 grams × 1.0027 = 50.0016 cm³.
Conclusion
Calibration of the burette is complete with a calculated volume of 50.0016 cm³.
📄